CVE & CISA-KEV Catalog

CVE-2014-5406

HIGH
7.6
CVSS v2
NVD

Description

The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.

How to fix

No published remediation has been found for this vulnerability's affected products yet.

Mitigation guidance may be in the linked vendor advisories in the References section below.

CVSS v2 Vector

No CVSS vector data available.

Exploit Intelligence

1.24%probability of exploitation in 30 days
66thpercentile

Moderate risk: more likely to be exploited than 66% of all known CVEs.

References

Embed a live status badge for CVE-2014-5406
CVE-2014-5406 severity badge

Markdown

[![CVE-2014-5406](https://tridentstack.com/cve/badge/CVE-2014-5406.svg)](https://tridentstack.com/cve/CVE-2014-5406)

HTML

<a href="https://tridentstack.com/cve/CVE-2014-5406"><img src="https://tridentstack.com/cve/badge/CVE-2014-5406.svg" alt="CVE-2014-5406"></a>

Find and fix vulnerabilities across your fleet

TridentStack Control continuously scans your Windows, macOS, and Linux fleet for known vulnerabilities, prioritizes them by severity and active exploitation, and patches them automatically.

Start free

This product uses NVD data but is not endorsed or certified by the NVD. EPSS scores courtesy of FIRST.org (https://www.first.org/epss). Source: CISA KEV Catalog. Data as of 2026-05-06.