Description

Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.

CVSS v3 Vector

Exploitability

Attack VectorNetwork
Attack ComplexityLow
Privileges RequiredNone
User InteractionNone
ScopeUnchanged

Impact

ConfidentialityHigh
IntegrityHigh
AvailabilityHigh

CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

Exploit Intelligence

0.50%probability of exploitation in 30 days
39thpercentile

Low risk: more likely to be exploited than 39% of all known CVEs.