CVE & CISA-KEV Catalog

CVE-2025-15653

MEDIUM
6.8
CVSS v3
NVD

Description

Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.

How to fix

No published remediation has been found for this vulnerability's affected products yet.

Mitigation guidance may be in the linked vendor advisories in the References section below.

CVSS v3 Vector

Exploitability

Attack VectorPhysical
Attack ComplexityLow
Privileges RequiredNone
User InteractionNone
ScopeUnchanged

Impact

ConfidentialityHigh
IntegrityHigh
AvailabilityHigh

CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

Exploit Intelligence

0.17%probability of exploitation in 30 days
7thpercentile

Low risk: more likely to be exploited than 7% of all known CVEs.

References

Embed a live status badge for CVE-2025-15653
CVE-2025-15653 severity badge

Markdown

[![CVE-2025-15653](https://tridentstack.com/cve/badge/CVE-2025-15653.svg)](https://tridentstack.com/cve/CVE-2025-15653)

HTML

<a href="https://tridentstack.com/cve/CVE-2025-15653"><img src="https://tridentstack.com/cve/badge/CVE-2025-15653.svg" alt="CVE-2025-15653"></a>

Find and fix vulnerabilities across your fleet

TridentStack Control continuously scans your Windows, macOS, and Linux fleet for known vulnerabilities, prioritizes them by severity and active exploitation, and patches them automatically.

Start free

This product uses NVD data but is not endorsed or certified by the NVD. EPSS scores courtesy of FIRST.org (https://www.first.org/epss). Source: CISA KEV Catalog. Data as of 2026-06-04.