MEDIUMDescription
Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.
How to fix
No published remediation has been found for this vulnerability's affected products yet.
Mitigation guidance may be in the linked vendor advisories in the References section below.
CVSS v3 Vector
Exploitability
Impact
CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:N/I:N/A:L