CVE-2015-3954
CRITICALEPSS 79th pctlDescription
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
How to fix
No published remediation has been found for this vulnerability's affected products yet.
Mitigation guidance may be in the linked vendor advisories in the References section below.
CVSS v3 Vector
Exploitability
Impact
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Exploit Intelligence
Elevated risk: more likely to be exploited than 79% of all known CVEs.
References
Embed a live status badge for CVE-2015-3954
Markdown
[](https://tridentstack.com/cve/CVE-2015-3954)HTML
<a href="https://tridentstack.com/cve/CVE-2015-3954"><img src="https://tridentstack.com/cve/badge/CVE-2015-3954.svg" alt="CVE-2015-3954"></a>Find and fix vulnerabilities across your fleet
TridentStack Control continuously scans your Windows, macOS, and Linux fleet for known vulnerabilities, prioritizes them by severity and active exploitation, and patches them automatically.
Start freeThis product uses NVD data but is not endorsed or certified by the NVD. EPSS scores courtesy of FIRST.org (https://www.first.org/epss). Source: CISA KEV Catalog. Data as of 2024-11-21.