CVE & CISA-KEV Catalog

CVE-2022-4992

HIGH
8.6
CVSS v3
NVD

Description

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions. Attackers can compromise network communications to modify device settings such as alarm states or alarm limits, or overwhelm the system with excessive network traffic causing the Cockpit or M540 to reboot and lose network functionality.

How to fix

No published remediation has been found for this vulnerability's affected products yet.

Mitigation guidance may be in the linked vendor advisories in the References section below.

CVSS v3 Vector

Exploitability

Attack VectorNetwork
Attack ComplexityLow
Privileges RequiredNone
User InteractionNone
ScopeUnchanged

Impact

ConfidentialityLow
IntegrityLow
AvailabilityHigh

CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H

Exploit Intelligence

0.16%probability of exploitation in 30 days
6thpercentile

Low risk: more likely to be exploited than 6% of all known CVEs.

References

Embed a live status badge for CVE-2022-4992
CVE-2022-4992 severity badge

Markdown

[![CVE-2022-4992](https://tridentstack.com/cve/badge/CVE-2022-4992.svg)](https://tridentstack.com/cve/CVE-2022-4992)

HTML

<a href="https://tridentstack.com/cve/CVE-2022-4992"><img src="https://tridentstack.com/cve/badge/CVE-2022-4992.svg" alt="CVE-2022-4992"></a>

Find and fix vulnerabilities across your fleet

TridentStack Control continuously scans your Windows, macOS, and Linux fleet for known vulnerabilities, prioritizes them by severity and active exploitation, and patches them automatically.

Start free

This product uses NVD data but is not endorsed or certified by the NVD. EPSS scores courtesy of FIRST.org (https://www.first.org/epss). Source: CISA KEV Catalog. Data as of 2026-06-04.