HIGHDescription
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.
How to fix
Remediation is compiled from vendor and distribution security advisories. Always confirm against the linked source for your exact version and platform.
CVSS v3 Vector
Exploitability
Impact
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N