CVE-2017-12714
MEDIUMDescription
Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted "RF wake-up" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker battery life. CVSS v3 base score: 5.3, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
How to fix
Remediation is compiled from vendor and distribution security advisories. Always confirm against the linked source for your exact version and platform.
CVSS v3 Vector
Exploitability
Impact
CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
Exploit Intelligence
Moderate risk: more likely to be exploited than 47% of all known CVEs.
References
Embed a live status badge for CVE-2017-12714
Markdown
[](https://tridentstack.com/cve/CVE-2017-12714)HTML
<a href="https://tridentstack.com/cve/CVE-2017-12714"><img src="https://tridentstack.com/cve/badge/CVE-2017-12714.svg" alt="CVE-2017-12714"></a>Find and fix vulnerabilities across your fleet
TridentStack Control continuously scans your Windows, macOS, and Linux fleet for known vulnerabilities, prioritizes them by severity and active exploitation, and patches them automatically.
Start freeThis product uses NVD data but is not endorsed or certified by the NVD. EPSS scores courtesy of FIRST.org (https://www.first.org/epss). Source: CISA KEV Catalog. Data as of 2024-11-21.